Table 3-2: Alarm Relay Contact Pins - Teledyne 2000RS Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

TCD Analyzer
Threshold Alarm 1:
 Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is below
threshold).
 Can be configured as failsafe or non-failsafe.
 Can be configured as latching or non-latching.
 Can be configured out (defeated).
Threshold Alarm 2:
 Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is below
threshold).
 Can be configured as failsafe or non-failsafe.
 Can be configured as latching or non-latching.
 Can be configured out (defeated).
System Alarm:
 Actuates when DC power supplied to circuits is unacceptable
in one or more parameters. Permanently configured as failsafe
and latching. Cannot be defeated. Actuates if self-test fails.
 Alarm 1 and Alarm 2 are configured from the Alarm
configuration screen.
Further detail can be found in Chapter 4 Section 4-7.

Table 3-2: Alarm Relay Contact Pins

Pin
45
Threshold Alarm 1, normally closed contact
28
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
Threshold Alarm 2, normally closed contact
44
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact