Cleaning and Maintenance

WARNING: Failure to clean Sample Measurement
Surface and Sample Well completely as described may
result in unreliable measurements or error messages.
See error message table.
Following each test, Refractix must be thoroughly
cleaned and dried as described below.
Use a clean, dry, residue-free wipe to remove sample
1
from the Sample Measurement Surface and Sample
Well. A cotton swab, alcohol wipe, or delicate task
wipe can also be used.
Rinse the Sample Measurement Surface and Sample
2
Well with distilled water, deionized water, or 70% IPA
(Isopropyl alcohol).
Dry the Sample Measurement Surface and Sample
3
Well with a clean, residue-free wipe.
Check surfaces to be sure that no residual haze is
4
present. If residue is present, re-clean surfaces.
CAUTION: Never immerse device in water.
The silicone protective
case can be removed
for cleaning. The device
housing and display can
be cleaned as well.
The following cleaning
agents are compatible
with Refractix and the
silicone protective case:
• Soap and water
• 70% IPA
(Isopropyl alcohol)
• Methanol
• 7.5% hydrogen peroxide
solution
• 5,000 PPM sodium hypochlorite
solution (Clorox, CA)
• Sani-Cloth® Prime Germicidal Wipes
(Didecyl Dimethyl Ammonium Chloride,
isopropyl alcohol, and ethyl alcohol)
• CaviCide®
13910000-101 Rev. A • 13