4. Operation

4.2 Setting RPM/RCF

1. Press the Setup/Enter button.
2. Press the Left or Right button until the RPM/RCF display(rpm) flashes.
▶ RPM setting mode: RPM(rpm) display flashes.
▶ RCF setting mode: RCF(xg) display flashes.
3. Press Up or Down button until the required RPM/RCF is displayed.
▶ RPM/RCF setting interval: 10 rpm/10 xg increase/decrease
▶ Press and hold the button to increase the setting interval by 100rpm/xg
4. Press the Setup/Enter button.

4.3 Setting Run Time

1. Press the Setup/Enter button.
2. Press the Left or Right button until the Time display(0' 00'') flashes.
▶ Setting run time in hours: 00' flashes
▶ Setting run time in minutes: 00'' flashes
3. Press Up or Down button until the required run time is displayed.
4. Press the Setup/Enter button.