3. Installation

3.5 Removing the Rotor

• Since this product is equipped with a rotor, no additional installation work is
• Since this product is equipped with a rotor, no additional installation work is
• Since this product is equipped with a rotor, no additional installation work is
required.
required.
required.
• Use only rotors produced by Hanil Scientific Inc.
• Use only rotors produced by Hanil Scientific Inc.
• Use only rotors produced by Hanil Scientific Inc.
• Do not use rotors that are corroded or have mechanical defects.
• Do not use rotors that are corroded or have mechanical defects.
• Do not use rotors that are corroded or have mechanical defects.
• Be sure to install the rotor lead before use.
• Be sure to install the rotor lead before use.
• Be sure to install the rotor lead before use.
When removing or replacing the rotor, follow the procedure below.
When removing or replacing the rotor, follow the procedure below.
When removing or replacing the rotor, follow the procedure below.
1. Remove the Rotor lid by gently pulling the rotor lid fastner upwards.
2. Hold the rotor with one hand and turn the rotor fixing nut counter - clockwise
using a wrench to loosen.
3. Pull the rotor upwards out of the motor shaft gently using both hands