3. Installation

3.1 Packing Inspection

• Check packing conditions carefully, before unpacking.
• Contact Hanil Scientific Inc. immediately if damages found.
• Check the delivery for completeness.
▶ You can get contact details on the packing box or manual.
3.2 Installation
Installation on hard and flat ground.
• Centrifuge should be installed on a hard and flat place.
• If the centrifuge is installed in an inclined place, the shaft may be bent
due to the weight of the rotor.
Good ventailation.
• For air circulation and safety, maintain a 30cm"clearance envelope"
around the centrifuge while the rotor is spinning.
Constant temperature/humidity
• Centrifuge equipped with the sensitive electronic software which is
fragile with humidity and temperature.
• Must avoid direct ray or heater and be put in the ambience of controlled
temp.and air.
Avoid the corrosive gas
• Install the centrifuge in a place where corrosive gas is not generated.
• Sulfur dioxide gas and chlorine gas may cause corrosion.
Leveling
• The shaft should be put exactly vertical on the hotizontaliy flat ground by
the leveling tool.