Advertisement

Quick Links

Operation Manual
M 13
Micro Centrifuge
Date of purchase
Serial No.

Advertisement

Table of Contents
loading

Summary of Contents for Hanil M 13

  • Page 1 Operation Manual M 13 Micro Centrifuge Date of purchase Serial No.
  • Page 2 Research Use Only Copyrightⓒ 2020 Hanil Scientific Inc. All rights reserved. Contact Us If you have any questions, contact Hanil Scientific Inc. or place of purchase. techsupport@ihanil.com / sales@ihanil.com UM-M13(E)(Rev.1), 20210122...
  • Page 3 Index 1. General Considerations -------------------------------------------------------- 4 1.1 Safety - 4 1.2 Transport & Storage - 5 1.3 Safety label - 5 1.4 Electric safety information - 5 2. Product Description ------------------------------------------------------------- 6 2.1 Appearance - 6 2.2 Delivery Package - 6 2.3 Technical Specifications - 7 3.
  • Page 4: General Considerations

    2. Check the voltage to be used, before connecting the centrifuge to the power source. 3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc. Hanil Scientific Inc. is not responsible for damages of the device and accidents caused by using parts and accessories not recommended.
  • Page 5: Usage Condition

    5. Do not block vents. 6. If you have the following emergencies, shut off the power supply and unplug the power cord from outlet and contact your place of purchase or Hanil Scientific Inc. - Unusual noises or smell from the device.
  • Page 6: Product Description

    ② Optional products [ See Page 16 (7. Rotors & Acc.)] - (A0.2-32) 32 x PCR tube or 4 x strip angle rotor - M 13 main body - (A 5.0 - 6) 6 x 5mL fixed angle rotor - Powercable - (A 2.0 - 12) 12 x 2mL fixed angle rotor...
  • Page 7: Technical Specifications

    2. Product Description Technical Specifications Max. RPM - (rotor) 13,500 RPM - (A2.0 - 12) 6,000RPM - (A0.2 - 32), (A5.0 - 6) Max. RCF 12,032 xg Time ≤ 15 min, Pulse Acc. time to max. speed ≤15 sec Dec. time from max. speed ≤8 sec Tube Capacity 1.5/2.0ml x 12...
  • Page 8: Installation

    3.1 Packing Inspection · Check packing conditions very carefully before unpacking. · Contact Hanil Scientific Inc. immediately if damages found. · Check the delivery for completeness. · You can get contact details on packing box or the bottom of the manual.
  • Page 9: Connect Power

    3. Installation 3.3 Connect Power · Connect the device to voltage sources which correspond to the electrical requirements on the label attached to the device. · Use sockets with a protective earth conductor and provided power cord. 1. Connect the power cable to the power socket on the back side of the device and plug the cord into the power outlet.
  • Page 10 3.5 Loading and Removal of the Rotor · This device is equipped with a rotor . · Only use rotors, parts, and accessories provided or validated by Hanil Sciedntific Inc. · Do not use defective rotor. Follow the steps below to remove or replace the rotor.
  • Page 11: Loading Tubes

    3. Installation 3.6 Loading tubes · Always use the same type of tube . · Tubes should be loaded symmetrically . · Do not exceed the maximum rated speed of the tube . · Same volume and density of sample should be put on tubes . ·...
  • Page 12: Operation

    4. Operation 4.1 Control Panel ① ② ④ ③ Button Description ① RPM setting knob Rotate clockwise or counter-clockwise to set the RPM value. Rotate clockwise or counter-clockwise to set the run time. ② Time setting knob (To use pulse function, set the time to '0' and press the start key.) ③...
  • Page 13: Setting Rpm

    4. Operation 4.2 Setting RPM [A2.0-12 rotor] 1. Rotate the 'RPM setting knob' clockwise or counter-clockwise to set the RPM value. ▶ 0~13,500 RPM [A0.2-32 rotor/A5.0-6 rotor] 1. Press the 6K RUN button to start operation at 6,000 RPM ▶ Do not exceed 6,000 RPM. 4.3 Setting Run Time 1.
  • Page 14: Maintenance

    2. Rotate the shaft with your hand to make sure it rotates smoothly. 3. Use the stopwatch to check the time setting for accuracy 4. If you find any damages, do not use the device. Contact Hanil Scientific Inc. 5. Be careful not to get any foreign substances on the sensor.
  • Page 15: Troubleshooting

    6. Troubleshooting 6.1 General Errors Problem Recommended Action Power failure Check the power cord connection. When opening the lid, error sound activated. Error sound doesn't stop Press the 6K Run button to stop the error sound. Lid cannot be opened Press the ‘Lid open button’. Remove the dirt on the lid latch and close the lid again.
  • Page 16: Rotors And Accessories

    7. Rotors and Accessories Tube Tube capacity 0.2 ml 0.5 ml 1.5/2.0 ml Adaptor None Fixed Angle Rotor, A2.0-12 Cat. No. TR0.2 TR0.5 Adaptor bore 12 x 1 . 5 / 2 . 0 ml 6 x 21 8 x 37 (Фx L, mm) Radius (mm) Hole diameter ( m m) : 11.5...
  • Page 17 MEMO...
  • Page 20 RECYCLABLE Doc.No. : UM-M13(E)(Rev.0)

Table of Contents