4.3 Loading/Unloading a rotor

•Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
•Do not use scatched or cracked rotor in use.
Loading
① Clean the motor shaft and the chamber.
- When loading the rotor, the temperature should be 10 to 30 degrees.
② Load the rotor vertically onto the motor shaft.
③ Insert the wrench into the center hole of the rotor
and turn it clockwise to fasten the rotor bolt.
④ Check the rotor to make sure it loaded firmly.
Unloading
① To remove the rotor insert the wrench into the center hole of the rotor and turn it
counterclockwise to loosen rotor bolt.