Section 4.0 Seismic

Note: Table related information is only applicable for Optima XR648.
4.1

Overview

Seismic requirements are determined and specified by the hospital/Design Professional of record
and may require approval by the specific state or country agency.
Seismic attachment hardware shown on seismic calculations may differ from hardware supplied
with system. Any additional hardware that is required will be the responsibility of the institution
and/or their contractor. Contact your local GE representative to obtain seismic calculations.
Seismic calculations are per California Building Code (CBC) and International Building Code (IBC).
4.2

Calculations

Seismic calculations can be obtained for the following:
• Overhead Tube Suspension (OTS)
• System Cabinet (SKL)
• Table (TBL)
• Manual Wall Stand (WLS)
• Standard Wall Stand (WLS)
• Extended Wall Stand (WLS)
• Detector BIN
Page 90
Section 4.0 - Seismic
Optima XR642/XR648 Pre-Installation