  • Page 1 Operating Instructions ATMOS LC 27 ATMOS LC 27 Battery English GA1GB.310301.0 2019-08 Index 02...
  • Page 2: Table Of Contents

    Removing the bacterial filter cartridge ....... . . 20 Further information, accessories, consumables and spare parts are available from: ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Straße 16...
  • Page 3 ATMOS LC 27 Battery ........
  • Page 4: Introduction

    § the device for the first time. This will help you to avoid potentially dangerous situations. The products ATMOS LC 27 and ATMOS LC 27 Battery bear CE marking CE 0124 according to the EC Directive of the council for medical products 93/42/EEC and meet the basic requirements of appendix I of the directive.
  • Page 5: Explanation Of Pictures And Symbols

    1.2 Explanation of pictures and symbols In the operating instructions DANGER Warning of a danger that will result in immediate fatal or serious injury. Observe the necessary measures. WARNING Warning of a danger that can cause fatal or serious injury. Observe the necessary measures.
  • Page 6 Serial number Order number European Article Number LOT number Expiry date Protected against ingress of: IP21 • Solid foreign particles Ø ≥12.5 mm • Dripping water from above Application part type BF Professional disposal Not for reuse (designed for repeated use for a single patient) Not sterile STERILE Connection for suction hose/patient...
  • Page 7: Intended Use

    Dripping water from above when the housing is tilled up to 15° Safety test seal 1.3 Intended use Name: ATMOS LC 27 | ATMOS LC 27 Battery Main function: Temporary and spontaneous suction of suction material (including secretions, blood, serum fluids, food particles) from the oral cavity, pharynx and bronchial system.
  • Page 8 Application organ: • Upper respiratory tract (nose, nasal cavity, pharynx) • Lower respiratory tract (larynx, trachea, bronchial system) Application time: Temporary use on the patient (< 60 min.) Area of application: • Home (homecare sector) • Outpatient and hospital care Criteria for patient selection: Patients who will benefit from upper/lower respiratory tract suction Indications:...
  • Page 9: Function

    The vacuum setting allows you to infinitely adjust the final vacuum and thus the suction capacity. The vacuum gauge shows the set vacuum. The ATMOS LC 27 and the ATMOS LC 27 Battery cannot be operated when in motion and therefore need to be kept stationary when operated. Single-patient secretion canister (DDS): The single-patient secretion canister is fastened to the side of the device and connects directly to the suction unit by means of the Direct-Docking-System (DDS).
  • Page 10: Scope Of Delivery

    1 x ATMOS mains power unit 011.1363.0 1 x operating instructions GA1GB.310301.0 Not included in the scope of delivery: • ATMOS LC 27 bag (REF 321.0350.0) • Car power cable (REF 313.0436.0) • ATMOS LC 27 / DDS consumables set, 10 pcs. (321.0340.0) Introduction...
  • Page 11: Transport And Storage

    If damage occurs during transport: 1. Document and report the transport damage. 2. Fill in the form QD 434 “delivery complaint/return shipment”. 3. Send the device to ATMOS (chapter “6.2 Sending in the device” on page 30). Ambient conditions for transport and storage: •...
  • Page 12: Notices For Your Safety

    2.0 Notices for your safety The safety of the ATMOS LC 27 and the ATMOS LC 27 Battery complies with all the recognized rules of technology and the guidelines of the Medical Products Law. 2.1 General safety instructions Make yourself familiar with the device in good time, so that you are capable of operating it when it is needed.
  • Page 13 In this case clean the device and send it in to ATMOS for repair. • The ATMOS LC 27 and ATMOS LC 27 Battery are not suitable for sterilisation. • Only use the mains power unit in dry surroundings.
  • Page 14: Avoiding Damage To The Device

    • Ensure that no liquid penetrates the device. In case that liquid has penetrated the device it may no longer be operated. In this case clean the device and send it in to ATMOS for repair. • Always place the device on firm, level surface. The device must always be in a vertical position, when you use it.
  • Page 15: Setting Up And Starting Up

    3.0 Setting up and starting up Please observe that insufficient battery charge can result in damage to the battery. 1. Fully charge the battery for the ATMOS LC 27 Battery before using the device. 3.1 Device overview 3.1.1 Front/rear view (with single-patient secretion canister system) Front view  Control panel  ...
  • Page 16: Control Panel

     Operation status display  The battery charge level of the ATMOS LC 27 Battery must be visually monitored by checking the indicator.  If only the red LED lights up, although the battery has been loaded for a longer period, the battery is defective.
  • Page 17: Charging The Battery

    8. Replace a defective mains power unit immediately. 9. ATMOS LC 27 Battery: Charge the battery fully, see chapter “3.3 Charging the battery” on page 17. 10. Operate the device only with the bacterial filter cartridge inserted and with the single-patient secretion canister system connected. The device is delivered with the bacterial filter cartridge and the single-patient secretion canister system already inserted and connected. Always keep an extra bacterial filter cartridge on hand.
  • Page 18: Charging Using The Mains Power Unit

    3.4 Charging using the mains power unit 1. Plug the mains power unit into the rear side of the device. 2. Plug the mains power unit into the power outlet. The device will be fully charged. » 3.5 Charging using the car power cable 1.
  • Page 19  Recommendation: Before beginning suction, fill a little water (around 10 - 20 ml) into the single-patient secretion canister system. This will prevent secretions from drying into crusts. 1. Place the canister lid onto the single- patient secretion canister. 2. Hold the single-patient secretion canister in place with one hand and screw on the single-patient secretion canister lid by turning it clockwise with a quarter turn.
  • Page 20: Notes On The Dds Bacterial Filter Cartridge

    3.6.2 Notes on the DDS bacterial filter cartridge  The DDS bacterial filter cartridge is located between the device and the single-patient secretion canister system. It consists of a hydrophobic bacterial filter and the blue holder with white seal. The bacterial filter cartridge effectively protects the device from oversuction and contamination.  The bacterial filter cartridge is designed for use on a single patient.
  • Page 21: Connecting The Suction Hose With Dds Adapter And Fingertip

    3.6.5 Connecting the suction hose with DDS adapter and fingertip 1. Insert the blue adapter on the suction hose into the opening on the single-patient secretion canister lid. 2. Turn the blue adapter clockwise with a quarter turn. 3. Now, check that the suction hose is connected firmly to the single-patient secretion canister system.
  • Page 22: Operation

    Air pressure: 700...1060 hPa 4.1 Switching on the device  The device should only be left on for as long as you need it. This will prolong the battery life (for ATMOS LC 27 Battery). 1. Press the on/off button to switch on the device. The pump starts. »...
  • Page 23: Suction

    3. Turn the vacuum adjustment to the right until the vacuum gauge shows the desired vacuum. 4. To reduce the vacuum, turn the vacuum adjustment to the left.  The vacuum adjustment shows the setting range directly (from - to +). ...
  • Page 24: Removing The Single-Patient Secretion Canister System

    4.5.1 Removing the single-patient secretion canister system 1. Remove the single-patient secretion canister system by pressing and holding the release button on the control panel and taking the single-patient secretion canister system out to the side. 2. Open the single-patient secretion canister lid by turning the lid anti-clockwise with a quarter turn.
  • Page 25: Cleaning And Disinfection

    5.0 Cleaning and disinfection 5.1 Safety instructions for reprocessing 5.1.1 General safety instructions We recommend you to always document all maintenance work and part replacements in writing. It is the responsibility of the user to ensure that the demands for cleaning and disinfection are adhered to.
  • Page 26: Preparing/Completing Reprocessing

    Unsuitable cleaning agents and disinfectants. Risk of damage to the product. • Do not use any process chemicals which contain the following ingredients on plastic parts • Chloramide or phenol derivatives • Do not use any process chemicals which contain the following ingredients on stainless steel •...
  • Page 27: Select Process Chemicals

    Surface Remarks Release button As per agent manufacturer’s instructions Control panel As per agent manufacturer’s instructions 5.3.2 Select process chemicals Observe the process chemical manufacturer’s specifications. If heavily soiled, clean the device surface directly with a cloth moistened with clean water. Agent (manufacturer) Active ingredients in 100 g Cleaning Mikrozid®...
  • Page 28: Wipe Cleaning

    5.3.4 Wipe cleaning 1. Disconnect the device from the mains power supply. 2. Clean the surface evenly with a suitable cloth and suitable cleaning agent, see chapter “5.3.1 Overview” on page 26. Pay particular attention to areas which are difficult to access. »...
  • Page 29 Suction hose with DDS adapter and fingertip/suction hose • Rinse out the suction hose with clean water after every use. • When using the device for only a single patient, replace the suction hose every 4 weeks. • If heavily soiled with clogged or crusted secretion, replace the suction hose immediately.
  • Page 30: Maintenance And Service

    Send the product in to ATMOS or your dealer. 6.3 Handling batteries The battery in the ATMOS LC 27 Battery is a wear part and is therefore not covered by the general warranty. There is a function guarantee of 6 months.
  • Page 31: Troubleshooting

    7.0 Troubleshooting The ATMOS LC 27 or ATMOS LC 27 Battery has been subjected to thorough quality inspection at the factory. Nevertheless, if a problem may occur you can possibly solve it yourself. Recharging and battery status Error symptom Possible cause...
  • Page 32 Vacuum and suction capacity Error symptom Possible cause Remedy Vacuum is not built Battery is discharged or Charge the battery or send the up or cannot be defective. device in for repair. reached. Hoses or single-patient Check the single-patient secretion secretion canister lid are canister lid and hoses for tight fit.
  • Page 33: Accessories And Consumables

    ATMOS LC 27 bag 321.0350.0 Car power cable 313.0436.0 ATMOS mains power unit 011.1363.0 8.2 Consumables Name ATMOS LC 27 / DDS consumables set, 10 pcs. 321.0340.0 Single-patient secretion canister system 321.0300.0 DDS bacterial filter cartridge, 10 pcs. 321.0027.0 Suction hose with DDS adapter and fingertip, 10 pcs. 321.0100.0 Suction hose, silicone, Ø 6mm, L = 1.3 m, 1 pcs.
  • Page 34: Disposal/Recycling

    Do not dispose of devices or batteries in domestic waste. 1. Clean and disinfect the device. 2. In Germany: Send in the device to ATMOS or your specialized dealer. They will dispose of the device professionally. 3. In other countries: Dispose of the device professionally and according to country- specific laws and regulations.
  • Page 35: Technical Data

    10.0 Technical data 10.1 ATMOS LC 27 Voltage 100 - 240 V~ ± 10%; 50/60 Hz (13.4 VDC mains power unit) Current consumption Max. 1 A Power consumption Max. 37 W Mains power unit Manufacturer: GlobTek, Inc. Model: GTM46402-3713.4 Suction capacity (on pump, during mains 27 l/min ±...
  • Page 36: Atmos Lc 27 Battery

    Basic UDI device identifier 04250365190830 ATMOS LC 27 / 04250365190861 ATMOS LC 27 / DDS Universal ID No. (REF) 321.1000.0 ATMOS LC 27 / DDS 321.1300.0 ATMOS LC 27 / DDS Universal 10.2 ATMOS LC 27 Battery Voltage 100 - 240 V~ ± 10%; 50/60 Hz (13.4 VDC mains power unit)
  • Page 37 MDA 0318 Basic UDI device identifier 04250365190878 ATMOS LC 27 Battery / DDS 04250365190908 ATMOS LC 27 Battery / DDS Universal ID No. (REF) 321.2000.0 ATMOS LC 27 Battery / 321.2300.0 ATMOS LC 27 Battery / DDS Universal * Data may differ depending on elevation above sea level, prevalent air pressure and air temperature. Technical data...
  • Page 38: Notes On Emc

     Medical electrical equipment is subject to special precautions with regard to EMC and must be installed acc. to following EMC notes. Guidance and manufacturer's declaration - ambient conditions The ATMOS LC 27 and ATMOS LC 27 Battery are suitable for operation in the following surroundings: • Homecare in any building.
  • Page 40 ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Straße 16 79853 Lenzkirch / Germany Phone: +49 7653 689-0 info@atmosmed.de www.atmosmed.com...