Medi-Vac Canister System - Atmos 318.0000.0 Operating Instructions Manual

Table of Contents

Advertisement

4. Connect the vacuum hose to the Serres
5. Check whether the foil of the Serres
external canister and the lid tightly fits to the Serres
6. Connect the disposable suction hose with the angle
7. Close the auxiliary air vent of the fingertip and close the front opening with your
thumb.
8. Switch on the device so that the pump builds up a vacuum.
The Serres
»
9. Insert the suction hose into the hose guide.
10. Wrap the suction hose onto the hose rewind.
11. Perform a function check: Chapter „6.2 Functional check" on page 39.
3.4.3 Medi-Vac
WARNING
Risk of infection by contaminated canister system and hoses.
Deadly diseases can be transmitted.
Only use Medi-Vac
Only use sterile packaged parts, when the packaging is undamaged.
No vacuum or vacuum is too low because of incorrect connection.
Patient can suffocate.
Please observe the operating instructions from the manufacturer of the Medi-Vac
canister system.
Removal
1. Remove the disposable suction hose from the hose guide.
2. Remove the disposable suction hose and the angle
bag.
22
Setting up and starting up
1
suction bag develops.
®
canister system
®
suction bags with integrated bacterial filter.
®
1
outer canister (grey angle
®
2
suction bag is fully inserted into the Serres
®
2
3
4
5
).
1
external canister.
®
to the Serres
suction bag.
®
2
from the Medi-Vac
suction
®
2
®
®