7.0 Eliminating errors

The ATMOS C 341 Battery was subjected to a thorough quality control in the factory.
Nevertheless, if a problem may occur you can possibly solve it yourself.
Recharging and battery status
Error indication
Possible cause
Device cannot be
The plug from the
recharged.
charging accessories is
poorly fitted.
Charging accessories
are defective.
Battery is not
connected properly.
Battery temperature
too high or too low.
Defective battery.
Defective fuse or
electronics.
LED on the power
Defective power
supply and
supply and recharging
recharging unit is
unit.
not illuminated.
The mains plug is
poorly fitted.
When recharging
Battery service life
the battery 100 %
is exhausted or the
cannot be achieved.
battery is defective.
The charging time
Wrong charging
can take up to 14
accessories.
hours.
Red LED of the
Defective battery.
battery status
indicator is lit up
although the battery
is fully charged.
Battery
Battery is not fitted
compartment cover
correctly.
cannot be closed.
Battery compartment
cover is installed
incorrectly.
Remedy
1. Check the connection to the mains
supply.
1. Exchange charging accessories.
1. Check the plug connections in the
battery compartment.
1. After long-term use: Let the device
cool down.
2. Extreme ambient temperature:
Position the unit where appropriate
to a cooler or warmer place.
1. Exchange the battery.
1. Send in the device for repair.
1. Exchange the power supply and
recharging unit.
1. Check the connection to the mains
supply.
1. Exchange the battery.
1. Only use the provided charging
accessories or an original spare part.
1. Exchange the battery.
1. Fit the battery correctly.
1. Mount the battery compartment
cover correctly according to
operating instructions.