Application time: Temporary use on the patient during respiratory tract ATMOS A 161 / 261 Aspirator / DDS ® suction (< 60 min.) or short-term use at ENT treatment (up to 30 days).
Scope of supply ● Prior to dispatch, this ATMOS device was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note). In addition to the basic device, the scope of delivery comprises the following parts:...
For your Safety ● The safety standard of the ATMOS A / C 161 / ® ● Prior to starting up the ATMOS A / C 161 / 261 ® 261 Aspirator corresponds with recognized me- Aspirator, read these operating instructions dical technical regulations and the directions of carefully.
Shove blind for covering the operating elements (ATMOS C 161 / 261 Aspirator) ® Vacuum adjustment Vacuum gauge (ATMOS C 161 / 261 Aspirator) ® Hose storage (fi g. 1, page 4) Suction hose storage (accessory) Switch On I / Off O Fig 2a.
General operation Prior to these notes please read the foregoing chapter of your respective version of the ATMOS A / C 161 / 261 Aspira- ® tor! Attention: Suction procedures in the ● Check the vacuum readout regularly! ● respiratory tract may only be ●...
Operation DDS ● The DDS bacterial fi lter / oversuction stop ● The device may not be operated is disposable. without DDS bacterial fi lter / over- suction stop! An exchange of the DDS bacterial fi lter is Therefore always hold ready a necessary before every change of patient.
Operation Receptal ® Holder for Receptal secretion ® canister Plug the screw thread of the holder from the top into the borehole at the bottom side of the device. Screw it together with the provided screw nut. Fig. 11. Assembling the Receptal ®...
Operation Medi-Vac and Serres ® ® Medi-Vac ® Holder for Medi-Vac secretion ® canister Plug the screw thread of the holder from the top into the borehole at the bottom side of the device. Screw it together with the provided screw nut. Assembling the Medi-Vac ®...
ATMOS partner or by a specialist, who is authorised by ATMOS, in compliance with Medical Devices Operator Ordinance, Medical Devices Act and BV-Med rules is required (please see chapter 9.2 reprocessing).
The device must be reprocessed by the manufacturer a damp (not wet) cloth and afterwards be disinfected or a certifi ed ATMOS partner if the device is oversucked. A with a surface disinfectant stated on page 14. contaminated suction device bears a risk for the patient as well ...
Non-ionic tensides (Application concentrate) Solvents, complexing agents Recommended disinfectants for surfaces Disinfectant Contents (in 100 g) Manufacturer ATMOS Green & Clean SK Alkyldimethylbenzylammoniumchloride <1 g Metasys, Rum (Austria) (application solution) Dialkyldimehtylammoniumchloride <1 g Alkyldimethylethylbenzylammoniumchloride <1 g Dismozon pur...
Cleaning and servicing plan for ATMOS A 161 / A 261 / C 161 / C 261 Aspirator ® start date: name of the item: serial number: cleaning cleaning/exchange performed cleaning cleaning exchange of exchange of exchange secretion canister of the suction hose, bacterial fi...
Page 16
ATMOS bacterial fi lter fi ed ATMOS partner or an ATMOS authorized, specialized dealer. The suction device may not be operated without bacterial fi lter! This cleaning and servicing plan as well as the relevant notes result Hose connection/fi...
● Maintenance or opening and repair of the ATMOS A / C 161 / 261 Aspirator (with the exception of the cleaning work the patient and the user and for maintaining a safe and described in these operating instructions) may only be carried reliable suction device.
10.0 Trouble-shooting Prior to dispatch, the ATMOS A / C 161 / 261 Aspirator was subjected to an extensive functional test. If, nevertheless, a failure ® should appear, you may possibly clear it yourself if you follow these notes Problem...
Support for Serres external canister 1 l ® rinsing canister set 11.2 Consumables Bacterial filter for DDS secretion canister, 10 pcs. 340.0054.0 Set of consumables for ATMOS A- and C-class DDS 313.0160.0 ® Device support General accessories Receptal external canister 1.5 l 310.0221.0...
A regular check of the condensate-controller on the rear side is necessary. Pull out the plastic plug at check the colour at the end of the hose. In case of discolouration/deposits a maintenance measure must be performed by a certifi ed ATMOS service partner! 13.2 Reprocessing (in perfect technical and hygienic condition)
Electromagnetic Environment - Guidance Harmonics IEC 61000-3-2 Class A The ATMOS A / C 161 / 261 Aspirator is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Page 24
5 Cycles for 5 Cycles ergy supply, it is recommended to supply the ATMOS A / C 161 / 261 70% U 70% U Aspirator from an uninterruptible (30 % Dip of the U (30 % Dip of the U current supply or a battery..
Page 25
TV stations cannot be predestined exactly. To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured field strength at the location where the ATMOS A / C 161 / 261 Aspirator is used...
Page 26
The ATMOS A / C 161 / 261 Aspirator is intended for use in electromagnetic environment in which ratiated distur- ® bances are controlled. The customer or user of the ATMOS C 161 Aspirator can help prevent electromagnetic ® interference by maintaining a minimum distance between portable and mobile RF Communications equipment and...