TECHNICAL DATA
7
Digital output
General
Supply current
Supply voltage effect
Ambient temperature effect
(amplifier only)
Response time (without
damping)
Pulse output
General
Specifications
Supply current
Supply voltage effect
Ambient temperature effect
(amplifier only)
Response time (without
damping)
80
Digital information is superimposed on a 4...20 mA signal at 1200 baud
10 mA DC maximum
No effect on accuracy
±0.01% of reading from -40...+80°C / -40...+176°F
0.5 seconds or the vortex shedding period for frequencies less than 2 Hz.
The pulse output is an externally powered 2-wire transistor switch type output.
This output can be configured using any applicable configuration device to select any
one of three types of pulse outputs: raw pulse, rate pulse and total pulse.
Raw pulse: This is the vortex shedding frequency directly passed through providing an
Raw pulse:
Raw pulse:
Raw pulse:
instantaneous, non dampened frequency output
Rate pulse: The frequency of this output is a 50% duty cycle pulse output with a
Rate pulse:
Rate pulse:
Rate pulse:
frequency range of 0...10, 0...100 or 0...1000 Hz, proportional to zero flow to the full
range flow rate/upper range value (URV).
Total pulse: The frequency of this output is also a 50% duty cycle pulse output that is
Total pulse:
Total pulse:
Total pulse:
configured to provide a pulse when a determined volumetric/totalized unit has flowed
through the meter.
Isolated 2-wire transistor switch
Applied voltage: 5...30 VDC
Maximum "ON" state voltage drop: 1.0 VDC
Maximum "ON" state current: 20 mA
Reverse polarity protected
Short circuit protected
Connectable to pull up or pull down counters
20 mA DC maximum
No effect on accuracy
±0.01% of reading from -40...+80°C / -40...+176°F
Raw pulse: Vortex shedding frequency
Rate or total pulse: 0.25 seconds or the vortex shedding period for frequencies less
than 2 Hz.
www.krohne.com
OPTISWIRL 5080
04/2020 - 4005615602 - MA OPTISWIRL 5080 R02 en