Quick Start
Quick Start
Quick Start
Quick Start

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OPTIFLUX 5000 and is the answer not in the manual?

Questions and answers

Summary of Contents for KROHNE OPTIFLUX 5000

  • Page 1 Quick Start Quick Start Quick Start Electromagnetic flowmeter in sandwich version The documentation is only complete when used in combination with the relevant documentation for the signal converter. © KROHNE 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 2: Table Of Contents

    3.3 Virtual reference for IFC 300 (C, W and F version) ............17 3.4 Connection diagrams ..................... 17 4 Technical data 4.1 Dimensions and weights ....................18 5 Notes www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 3: Safety Instructions

    If you need to return the device to the manufacturer or supplier, please fill out the form • contained in the manual and send it with the device. Unfortunately, the manufacturer cannot repair or inspect the device without the completed form. 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 4: Installation

    6 Grounding rings (optional) INFORMATION! Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives. www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 5: Device Description

    Product specific information and extensive product specification is available using PICK, the Product Information Center KROHNE web-tool. PICK can be found via the service menu button on the KROHNE.com website. The following versions are available: • Compact version (the signal converter is mounted directly on the flow sensor) •...
  • Page 6: Nameplates (Examples)

    2 Type designation of the flowmeter, serial number and manufacturing date 3 Calibration data 4 PED data 5 Disposal logo and CE sign with number(s) of notified body / bodies www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 7: Storage

    • Allen key (4 mm) • Small screwdriver • Wrench for cable glands • Wrench for wall mounting bracket (remote version only) • Torque wrench for installing flowmeter in pipeline 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 8: General Requirements

    IEC 60068-2-64. Avoid magnetic field! Keep at least 5 DN distance between electromagnetic flow sensors. • 2.7.1 Vibrations Figure 2-5: Avoid vibrations 2.7.2 Magnetic field Figure 2-6: Avoid magnetic fields www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 9: Installation Conditions

    INFORMATION! 2 dimensional bends occur in a vertical or or or or horizontal plane (X/Y) only, while 3 dimensional bends occur in both vertical and and horizontal plane (X/Y/Z). 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 10: T-Section

    1 ≥ 10 DN 2.8.4 Bends Figure 2-10: Installation in bending pipes (90°) Figure 2-11: Installation in bending pipes (45°) CAUTION! Avoid draining or partial filling of the flow sensor www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 11: Open Discharge

    Figure 2-12: Installation in front of an open discharge 2.8.6 Flange deviation CAUTION! Max. permissible deviation of pipe flange faces: ≤ 0.5 mm / 0.02" Figure 2-13: Flange deviation 2.8.7 Pump Figure 2-14: Installation behind a pump 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 12: Control Valve

    2.8.9 Air venting and vacuum forces Figure 2-16: Air venting 1 ≥ 5 m / 17 ft 2 Air ventilation point Figure 2-17: Vacuum 1 ≥ 5 m / 17 ft www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 13: Mounting Position

    INSTALLATION OPTIFLUX 5000 2.8.10 Mounting position Figure 2-18: Mounting position 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 14: Mounting

    150 lb 1/10...3" 300 lb CAUTION! Pressures at 20 C / 68 • ° ° • For higher temperatures, the pressure and temperature ratings are as per ASME B16.5. www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 15 150 lb 1/2"UNC x 142 1 1/2" 150 lb 1/2"UNC x 174 2" 150 lb 5/8"UNC x 215 3" 150 lb 5/8"UNC x 268 4" 150 lb 5/8"UNC x 318 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 16: Electrical Connections

    2 Metal pipelines with internal coating and non-conductive pipelines. Grounding with grounding rings! INFORMATION! For diameter DN10 / 3/8" and DN15 / ", grounding rings are integrated as standard in the ½ flow sensor construction. www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 17: Virtual Reference For Ifc 300 (C, W And F Version)

    • Signal cable: max. 50 m / 164 ft, type DS 3.4 Connection diagrams INFORMATION! For the connection diagrams, please refer to the documentation of the applicable signal converter. 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 18: Technical Data

    IFC 100 (45 IFC 100 (45 IFC 100 (45°) ) ) ) b = 161 mm / 6.3" c = 184 mm / 2.7" Total height = H + a www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 19 IFC 050 (10 IFC 050 (10 b = 157 mm / 6.18" c = 260 mm / 10.24" 1 The value may vary depending on the used cable glands. 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 20 Note that for other pressure ratings than mentioned, the dimensions may be different. • For full information on signal converter dimensions see relevant documentation. • www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 21 5.24 19.40 1 Total fitting length of flowmeter with integrated rings: dimension L + 2 x gasket thickness. 2 Total fitting length of flowmeter without rings: dimension L only. 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 22: Notes

    NOTES OPTIFLUX 5000 www.krohne.com 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en...
  • Page 23 NOTES OPTIFLUX 5000 06/2021 - 4004943903 - QS OPTIFLUX 5000 SW R06 en www.krohne.com...
  • Page 24 Engineering, commissioning, calibration, maintenance and training services Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.: +49 203 301 0 Fax: +49 203 301 10389 info@krohne.de The current list of all KROHNE contacts and addresses can be found at: www.krohne.com...