KROHNE OPTISWIRL 5080 Handbook page 18

Vortex flowmeter for high temperature applications
Hide thumbs Also See for OPTISWIRL 5080:
Table of Contents

Advertisement

INSTALLATION
3
Figure 3-2: Not recommended installation
1 Installing the device in a downstream pipe
2 Installing the device in front of an outlet
3 Installing the device in an upper pipe bend due to risk of gas bubbles forming
CAUTION!
Installing the device in a downstream pipe 1 or upstream pipe of an outlet 2,
there is a risk of partially filled pipes leading to inaccurate measurements.
Installing the device in an upper pipe bend 3, there is a risk of gas bubbles forming.
Gas bubbles can lead to pressure surges and inaccurate measurement.
18
www.krohne.com
OPTISWIRL 5080
04/2020 - 4005615602 - MA OPTISWIRL 5080 R02 en

Advertisement

Table of Contents
loading