Glass Breakage; Lifetime Of Rotors And Accessories - Hermle Z 207 H User Manual

Table of Contents

Advertisement

MAINTENANCE
Chemical Sterilization
Bottles, adapters and rotors may be treated, with the usual liquid disinfectants.
ATTENTION: Before applying any other cleaning resp. decontamination method other than
what was recommended by the manufacturer, contact the manufacturer to ensure that it will
not damage the unit or the rotor.
.

3.1.6 Glass Breakage

With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed
immediately from rotor, adapters and the rotor chamber itself. Fine glass splinters will scratch and
therefore damage the protective surface coating of a rotor. If glass splinters remain in the rotor
chamber, fine metal dust will build up, due to air circulation. This very fine, black metal dust will
severely pollute the rotor chamber, the rotor and the samples.
If necessary, replace the adapters, tubes and accessories, to avoid further damage. Check the rotor
bores regularly, for residue and damage.
ATTENTION: Please check the relevant specifications of the tubes with the manufacturer!

3.2 Lifetime of Rotors and Accessories

Rotors and rotor lid made of aluminum or stainless steel, have a operating time of max. 7 years
from first use.
Transparent rotor lids and caps, made of PC or PP, as well as rotors, tube racks and adapters of
PP, have a maximum operating time of up to 3 years, from first time use.
Conditions for the operating time:
Proper use, damage-free condition, recommended care.
Z207H_V1.22_eng
24
© Hermle Labortechnik GmbH

Advertisement

Table of Contents
loading

Table of Contents