pH value
Carrying out a QSC
control calibration
42
sor evaluation.
0
pH
7.007
01.02.2014
Info
1 QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal cali-
brations.
1.
Open the menu for measurement settings with <MENU/ENTER>.
2.
In the QSC menu, select Control calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.
3.
Follow the user guide.
The calibration is carried out like a normal three-point calibration. As
soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.
14
25.0
°C
USB output
08:00
MultiLab 4010-1
1
ba76141e03
03/2014
Need help?
Do you have a question about the YSI MultiLab 4010P-1 and is the answer not in the manual?
Questions and answers