Xylem YSI MultiLab 4010P-1 Operation Manual page 91

Table of Contents

Advertisement

MultiLab 4010-1
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter or IDS sensor
ba76141e03
03/2014
Cause
– Batteries almost empty
Cause
– Operating condition undefined
or EMC load unallowed
Cause
– E. g., a question by the service
department
Remedy
– Replace the batteries (see section
13.1 M
AINTENANCE
Remedy
– Processor reset:
Press the <ENTER> and <On/Off>
key simultaneously
Remedy
– Switch on the meter.
– Open the menu, <ENTER__> / Stor-
age & config / System / Service infor-
mation. The instrument data are
displayed.
or
– Connect the sensor.
Press softkey [<F1>Info<F1>] /
[More] The sensor data are displayed
(see section 4.1.5 S
18)
What to do if...
, page 86)
, page
ENSOR INFO
91