MultiLab 4010-1 Overview Overview MultiLab 4010-1 meter The compact, digital precision meter MultiLab 4010-1 enables you to carry out pH measurements, ORP measurements, conductivity measurements and dis- solved oxygen (D.O.) measurements quickly and reliably. The MultiLab 4010-1 provides the maximum degree of operating comfort, reli- ability and measuring certainty for all applications.
Overview MultiLab 4010-1 MultiLab 4010P-1 meter with integrated printer The integrated printer of the MultiLab 4010P-1 enables to document measure- ments according to GLP requirements. The information concerning the printer of the MultiLab 4010P-1 is given in a separate chapter (see section 12 P RINTER ONLY 4010P-1), page 83).
MultiLab 4010-1 Overview have quick-lock couplings with which to fix the sensors to the meter. Sensor data from IDS sensors transmit the following sensor data to the meter: IDS sensors SENSOR ID – Sensor name – Sensor series number ...
MultiLab 4010-1 Safety Safety Safety information 2.1.1 Safety information in the operating manual This operating manual provides important information on the safe operation of the meter. Read this operating manual thoroughly and make yourself familiar with the meter before putting it into operation or working with it. The operating manual must be kept in the vicinity of the meter so you can always find the infor- mation you need.
Safety MultiLab 4010-1 Safe operation 2.2.1 Authorized use The authorized use of the meter consists exclusively of the measurement of the pH, ORP, conductivity and dissolved oxygen in a laboratory environment. Only the operation and running of the meter according to the instructions and technical specifications given in this operating manual is authorized (see section 15 T , page 92).
MultiLab 4010-1 Commissioning Commissioning Scope of delivery Meter MultiLab 4010-1 / MultiLab 4010P-1 4 batteries 1.5 V Mignon type AA Power pack USB cable (A plug on mini B plug) Stand Stand holder Comprehensive operating manual ...
Commissioning MultiLab 4010-1 3.3.1 Inserting the batteries You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH). In order to charge the batteries, an external charging device is required. Open the battery compartment (1) on the underside of the meter.
MultiLab 4010-1 Commissioning 3.3.2 Connecting the power pack CAUTION The line voltage at the operating site must lie within the input voltage range of the original power pack (see section 15.2 , page 92). ENERAL DATA CAUTION Use original power packs only (see section 15.2 G ENERAL DATA page 92).
Operation MultiLab 4010-1 Operation General operating principles 4.1.1 Keypad In this operating manual, keys are indicated by brackets <..>. The key symbol (e.g. <MENU/ENTER>) generally indicates a short keystroke (press and release) in this operating manual. A long keystroke (press and hold for approx.
MultiLab 4010-1 Operation 4.1.2 Display Example (pH): AutoCal TEC HOLD AR 03.04.2013 USB output Info 08:00 1 Status information (sensor) 2 Measured value 3 Measured parameter 4 Continuous measurement control (CMC function) 5 Sensor symbol (calibration evaluation, calibration interval) 6 Measured temperature (with unit) 7 Status information (meter) 8 Softkeys and date + time 4.1.3 Status information (meter)
Operation MultiLab 4010-1 4.1.4 Instrument connectors 1 IDS sensors: (pH, ORP, conductivity, D.O.) 2 USB B (device) interface 3 Stirrer (ProOBOD) 4 Service interface 5 Power pack CAUTION Only connect sensors to the meter that cannot return any volt- ages or currents that are not allowed (> SELV and > current circuit with current limiting).
MultiLab 4010-1 Operation 7.007 25.0 °C 4110 03.04.2013 More B09250001 08:00 Display further sensor data (settings) with <F1>/[More]. 4110 B092500013 pH resolution 0.001 mV resolution Buffer Calibration interval Unit for slope mV/pH QSC: Software-Version V 1.00 03.04.2013 Back 08:00 Switching on the meter Switch the meter on with <On/Off>.
Operation MultiLab 4010-1 If the user administration function is activated, the Login dialog appears after the meter is switched on (see section 4.4 L OGIN WITH , page 20). A USER NAME The user administration function is not active in the delivery condi- tion.
Page 21
MultiLab 4010-1 Operation The login is done immediately if no password is required. If a sensor is connected the measured value display appears. If a password is required: Using <><>, select the menu item, Password and confirm with <MENU/ENTER>. The user specifies the password when he or she first logs in with a user name.
Operation MultiLab 4010-1 Navigation 4.5.1 Operating modes Operating Description mode Measuring The measurement data of the connected sensor are shown in the measured value display Calibration The course of a calibration with calibration information, functions and settings is displayed Storing in The meter stores measuring data automatically or manu- memory ally...
Page 23
MultiLab 4010-1 Operation System General Interface Clock function Service information Reset 03.04.2013 Back 08:00 Settings Settings are indicated by a colon. The current setting is displayed on the right-hand side. The setting mode is opened with <MENU/ENTER>. Subse- quently, the setting can be changed with <><> and <MENU/ENTER>. Example: General Language:...
Operation MultiLab 4010-1 Calibration record Calibration data storage Buffer: Single-point calibration: Calibration interval: Unit for slope: mV/pH ] 4.00 7.00 10.00 03.04.2013 Back 08:00 4.5.4 Example 1 on navigation: Setting the language Press the <On/Off> key. The measured value display appears. The instrument is in the measuring mode.
Page 25
MultiLab 4010-1 Operation System General Interface Clock function Service information Reset 03.04.2013 Back 08:00 Select the General submenu with <><>. The current selection is displayed with a frame. Open the General submenu with <MENU/ENTER>. General Language: English Audio signal: Illumination: Contrast: 50 % Shutoff time:...
Operation MultiLab 4010-1 4.5.5 Example 2 on navigation: Setting the date and time The meter has a clock with a date function. The date and time are indicated in the status line of the measured value display. When storing measured values and calibrating, the current date and time are automatically stored as well.
Page 27
MultiLab 4010-1 Operation Change and confirm the setting with <><> and <MENU/ENTER>. The time is set. If necessary, set the Date and Date format. The setting is made similarly to that of the time. To make further settings, switch to the next higher menu level with <F1>/ [Back].
pH value MultiLab 4010-1 pH value Measuring 5.1.1 Measuring the pH value NOTE When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. Connect the IDS pH sensor to the meter. The pH measuring window is displayed.
MultiLab 4010-1 pH value You can terminate the Stability control function and the HOLD func- tion with <AR> or <M> at any time. Using <MENU/ENTER>, activate the Stability control function manu- ally. The [AR] status indicator appears while the measured value is assessed as not stable.
pH value MultiLab 4010-1 pH calibration 5.2.1 Why calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines and stores the current values of the zero point and slope of the electrode.
Page 31
MultiLab 4010-1 pH value Immerse the sensor in the first buffer solution. Start the measurement with <MENU/ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes. Buffer 10.00 -180.0 24.8 °C AutoCal YSI 01.07.2011...
Page 32
pH value MultiLab 4010-1 7.000 Buffer 24.8 °C AutoCal YSI 01.07.2011 08:00 Wait for the measurement with stability control to be completed or ter- minate the stability control and take over the calibration value with <MENU/ENTER>. The calibration display for the next buffer appears (voltage display). If necessary, finish the calibration procedure as a two-point calibration with <M>.
MultiLab 4010-1 pH value Calibration is automatically completed after the last buffer of a buf- fer set has been measured. Then the calibration record is dis- played. The calibration line is determined by linear regression. 5.2.4 Carrying out manual calibration (ConCal) Make sure that in the sensor menu, Buffer menu, the ConCal buffer set is correctly selected (see section 9.1.1 S , page...
Page 34
pH value MultiLab 4010-1 7.000 Buffer 24.8 °C ConCal 01.07.2011 08:00 Wait for the measurement with stability control to be completed or ter- minate the stability control and take over the calibration value with <MENU/ENTER>. The pH value of the buffer solution is displayed. Buffer 7.000 ...
Page 35
MultiLab 4010-1 pH value Wait for the measurement with stability control to be completed or ter- minate the stability control and take over the calibration value with <MENU/ENTER>. The pH value of the buffer solution is displayed. Buffer 4.035 ...
pH value MultiLab 4010-1 Set the nominal buffer value for the measured temperature with <><>. Accept the calibration value with <MENU/ENTER>. The calibration display for the next buffer appears (voltage display). If necessary, use <M> to finish the calibration. The calibration record is displayed. Continue calibrating using the next buffer with <MENU/ENTER>.
MultiLab 4010-1 pH value 5.2.6 Calibration data The calibration record is automatically transmitted to the interface after calibrating. Displaying the The calibration record of the last calibration is to be found under the menu item, calibration data Calibration / Calibration record. To open it in the measured value display, press the <CAL__>...
Page 38
pH value MultiLab 4010-1 Calibration After calibrating, the meter automatically evaluates the calibration. The zero evaluation point and slope are evaluated separately. The worse evaluation of both is taken into account. The evaluation appears on the display and in the calibration record.
MultiLab 4010-1 pH value 5.2.7 Continuous measurement control (CMC function) The Continuous Measurement Control (CMC function) facilitates to evaluate the current measured value instantly and definitely. After each successful calibration the scale of the pH measuring range is dis- played in the measured value display. Here you can very clearly see whether or not the current measured value is in the calibrated part of the measuring range.
pH value MultiLab 4010-1 5.2.8 QSC function (sensor quality control) General information The QSC function (Quality Sensor Control) is a new sensor evaluation for on the QSC function digital IDS sensors. It evaluates the condition of an IDS pH sensor individually and with a very fine grading.
Page 41
MultiLab 4010-1 pH value Carry out the normal calibrations for your special measuring range with your usual standard solutions as previously done. As soon as the QSC function was enabled for an IDS sensor, it is not possible to return to the sensor evaluation with the sensor symbol for this sensor.
Page 42
pH value MultiLab 4010-1 sor evaluation. 7.007 25.0 °C 01.02.2014 Info USB output 08:00 1 QSC scale The double arrow on the QSC scale indicates the current sensor evaluation Carrying out a QSC A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably control calibration changed (after some normal calibrations).
MultiLab 4010-1 Measuring 6.1.1 Measuring the ORP NOTE When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. IDS ORP sensors are not calibrated. However, you can check IDS ORP sensors using a test solution.
MultiLab 4010-1 You can terminate the Stability control function and the HOLD func- tion with <AR> or <M> at any time. Using <MENU/ENTER>, activate the Stability control function manu- ally. The [AR] status indicator appears while the measured value is assessed as not stable.
MultiLab 4010-1 Dissolved oxygen Dissolved oxygen Measuring 7.1.1 Measuring D.O. Connect the IDS D.O. sensor to the meter. The D.O. measuring screen is displayed. If necessary, select the measured parameter with <M>. Check or calibrate the meter with the sensor. Immerse the IDS D.O.
Page 46
Dissolved oxygen MultiLab 4010-1 The DO % Saturation local function for the parameter oxygen sat- uration [%] is switched on or off in the menu for calibration and measurement settings (see section 9.3.1 S D.O. ETTINGS FOR ), page SORS MENU FOR MEASUREMENT AND CALIBRATION SETTINGS 63).
MultiLab 4010-1 Dissolved oxygen Criteria for a stable The Stability control function checks whether the measured values are stable measured value within the monitored time interval. Measured parame- Time interval Stability in the time interval D.O. concentration 20 seconds ∆ : better than 0.03 mg/l D.O.
Dissolved oxygen MultiLab 4010-1 7.2.4 1-point calibration Calibration in water vapor-saturated air (Cal) and calibration via Comparison meas. according to the Winkler method (Comp) are described jointly here. Connect the IDS D.O. sensor to the meter. Check the selected calibration type and selected number of calibration points in the sensor setting menu and, if necessary, change them.
MultiLab 4010-1 Dissolved oxygen Wait for the AutoRead measurement to be completed (audio signal). When calibrating in water vapor-saturated air (Cal): The calibration record is displayed and output to the interface. When calibrating via comparison measurement (Comp): Set the concentration value with <> <> and accept it with <ENTER>.
Page 50
Dissolved oxygen MultiLab 4010-1 0.09 mg/l 25.1 °C 760 mmHg 03.04.2013 08:00 Start the measurement with <MENU/ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes. Wait for the AutoRead measurement to be completed (audio signal). The display for the second calibration point is displayed.
MultiLab 4010-1 Dissolved oxygen Wait for the AutoRead measurement to be completed (audio signal). When calibrating in water vapor-saturated air (Cal): The calibration record is displayed and output to the interface. When calibrating via comparison measurement (Comp): Set the concentration value with <> <> and accept it with <ENTER>.
Page 52
Dissolved oxygen MultiLab 4010-1 Menu item Setting/ Description function Calibration / Displays the calibration records. Calibration data stor- age /Display Further options: Scroll through the calibration records with <><>. Output the displayed calibra- tion record to the interface with <F2>/[USB output].
MultiLab 4010-1 Conductivity Conductivity Measuring 8.1.1 Measuring the conductivity NOTE When connecting a grounded PC/printer, measurements cannot be performed in grounded media as the values would be incorrect. The USB interface is not galvanically isolated. Connect the conductivity sensor to the meter. The conductivity measuring window is displayed.
Page 54
Conductivity MultiLab 4010-1 ting for automatic Stability control (see section 9.5.3 A UTOMATIC TABILITY CON , page 69) in the System menu. TROL Freeze the measured value with <AR>. The [HOLD] status indicator is displayed. The HOLD function is active. You can terminate the Stability control function and the HOLD func- tion with <AR>...
MultiLab 4010-1 Conductivity 8.1.2 Measuring the temperature For reproducible conductivity measurements, it is essential to measure the temperature of the test sample. IDS sensors measure the temperature with a temperature sensor integrated in the IDS sensor. Temperature compensation The calculation of the temperature compensation is based on the preset refer- ence temperature, 20 °C or 25 °C.
Conductivity MultiLab 4010-1 Calibration 8.3.1 Why calibrate? Aging slightly changes the cell constant, e. g. due to coatings. As a result, an inexact measured value is displayed. The original characteristics of the cell can often be restored by cleaning the cell. Calibration determines the current value of the cell constant and stores this value in the meter.
MultiLab 4010-1 Conductivity Immerse the conductivity sensor in the control standard solution, 0.01 mol/l KCI. Start the measurement with <MENU/ENTER>. The measured value is checked for stability (stability control). The [AR] status indicator is displayed. The measured parameter flashes. Wait for the end of the measurement with stability control ([HOLD][AR]) status indicator or take over the calibrated value with <MENU/ENTER>.
Page 58
Conductivity MultiLab 4010-1 Menu item Setting/ Description function Calibration / Outputs the calibration data storage Calibration data to the interface storage / Output to RS232/USB Calibration After calibration, the meter automatically evaluates the current status of the cal- evaluation ibration. The evaluation appears on the display and in the calibration record. Display Calibration record Cell constant [cm...
MultiLab 4010-1 Settings Settings pH measurement settings 9.1.1 Settings for pH measurements Settings The settings are made in the menu for calibration and measurement settings of the pH/ORP measurement. To open the settings, display the required mea- sured parameter in the measured value display and press the <MENU/ ENTER>...
Settings MultiLab 4010-1 Menu item Possible Description setting pH resolution 0.001 Resolution of the pH display 0.01 mV resolution Resolution of the mV display Reset Resets all sensor settings to the delivery condition (see sec- tion 9.6.1 R , page 70). ESETTING THE MEASUREMENT SETTINGS 9.1.2 Buffer sets for calibration You can use the buffer sets quoted in the table for an automatic calibration.
Page 61
MultiLab 4010-1 Settings Buffer set * pH values Merck 5 * 4.010 25 °C 7.000 10.000 DIN 19267 1.090 25 °C 4.650 6.790 9.230 Mettler Toledo USA * 1.679 25 °C 4.003 7.002 10.013 Mettler Toledo EU * 1.995 25 °C 4.005 7.002 9.208...
MultiLab 4010-1 Settings Confirm the setting with <MENU/ENTER>. Quit the menu with <M>. ORP measurement settings 9.2.1 Settings for ORP measurements The settings are made in the menu for measuring settings of the ORP measure- ment. To open the settings, display the required measured parameter in the measured value display and press the <MENU/ENTER>...
Page 64
Settings MultiLab 4010-1 Menu item Possible Description setting Calibration / 1-point calibration Calibration data storage / Number 2-point calibration of cal. points Calibration /Cali- 1 ... 180 ... Calibration interval for the IDS D.O. bration interval 999 d sensor (in days). The meter reminds you to calibrate regularly by the flashing sensor sym- bol in the measuring screen.
MultiLab 4010-1 Settings 9.3.2 Enter Sensor cap coefficients The values of the coefficients are provided with the sensor cap. Change the digit of the highlighted position with <><>. Go to the next position with <><>. Confirm with<ENTER> when a coefficient is completely entered. 9.3.3 DO % Saturation local Irrespective of the height or air pressure, the calibration value is set to 100%.
Settings MultiLab 4010-1 Cond measurement settings 9.4.1 Settings for IDS conductivity sensors Settings The settings are made in the menu for the measured parameter, conductivity. To open the settings, display the required measured parameter in the mea- sured value display and press the <MENU/ENTER> key. After completing the settings, switch to the measured value display with <M>.
Page 67
MultiLab 4010-1 Settings Menu item Possible Description setting Temp. comp. (TC) / 0.000 ... Coefficient of the linear temperature Linear coeff. 2.000 ... compensation. 3.000 %/K This menu item is only available when the linear temperature compensation is set. Temp. comp. (TC) / 20 °C Reference temperature Reference temp.
Settings MultiLab 4010-1 Sensor-independent settings 9.5.1 System To open the Storage & config menu, press the <MENU/ENTER__> key in the measured value display. After completing the settings, switch to the measured value display with <M>. Default settings are printed in bold. Menu item Setting Description...
MultiLab 4010-1 Settings Menu item Setting Description System /Reset Resets the system settings to the default values. For details, see section 9.6.2 R ESETTING , page 72 THE SYSTEM SETTINGS 9.5.2 Data storage This menu contains all functions to display, edit and erase stored measured values.
Settings MultiLab 4010-1 You can also generally switch on the display illumination (see section 9.5.1 , page 68). YSTEM Reset You can reset (initialize) all sensor settings and sensor-independent settings separately from each other. 9.6.1 Resetting the measurement settings The calibration data are reset to the default settings together with the measuring parameters.
Page 71
MultiLab 4010-1 Settings The following settings for ORP measurements are reset to the default settings with the Reset function: Setting Default settings mV resolution Man. temperature 25 °C The sensor settings are reset under the Reset menu item in the menu for cali- bration and measurement settings.
Settings MultiLab 4010-1 The sensor settings are reset under the Reset menu item in the menu for cali- bration and measurement settings. To open the settings, display the required measured parameter in the measured value display and press the <MENU/ ENTER>...
MultiLab 4010-1 Data storage Data storage You can save measured values (datasets) to the data storage: Manual data storage (see section 10.1 M , page 73) ANUAL DATA STORAGE Automatic data storage at intervals (see section 10.2 A UTOMATIC DATA STOR , page 74) AGE AT INTERVALS...
Data storage MultiLab 4010-1 10.2 Automatic data storage at intervals The data storage interval (Interval) determines the time interval between auto- matic data storage processes. Each data storage process transmits the current dataset to the USB interface. Configuring the Press the <STO__> key. automatic data The menu for automatic data storage appears.
Page 75
MultiLab 4010-1 Data storage Starting the To start the automatic data storage function, select continue with <><> and automatic data confirm with <MENU/ENTER>. The meter switches to the measured value dis- storage function play. 7.007 25.0 °C 0d03h00min 03.04.2013 08:00 1 Remaining storage duration 2 Graphical display of the storage duration The active automatic data storage function can be recognized by the progress...
Data storage MultiLab 4010-1 10.3 Measured value data storage 10.3.1 Editing the measured value data storage The contents of the manual or automatic measurement data storage can be shown on the display. Each of the measurement data storages has a function to erase the entire contents.
MultiLab 4010-1 Data storage Display Manual data storage 3 of 64 representation 03.04.2013 07:43:33 ID number: 1 of a dataset 4110 B20234008565 pH 7.000 24.8 °C AR Sensor: +++ 03.04.2013 USB output Back 08:00 Representation of a 03.04.2013 07:43:33 dataset MultiLab 4010-1 (USB output) Ser.
Data storage MultiLab 4010-1 ID number Measured value of the connected sensor Measured temperature value of the connected sensor AutoRead info: AR appears with the measured value if the AutoRead crite- rion was met while storing (stable measured value). Otherwise, the AR dis- play is missing.
MultiLab 4010-1 Transmitting data (USB interfaces) Transmitting data (USB interfaces) 11.1 Outputting current measurement data Output the current measurement data to the USB-B interface with <F2>[USB output]. 11.2 Transmitting data (to a PC) The meter has a USB-B interface (USB Device) e.g. to connect a PC. Via the USB-B interface (USB Device) you can transmit data to a PC or printer and update the meter software.
Transmitting data (USB interfaces) MultiLab 4010-1 11.4 Options for data transmission to a PC Via the USB-B interface you can transmit data to a PC. The following table shows which data are transmitted to the interface in which way: Data Control Operation / description Current...
MultiLab 4010-1 Transmitting data (USB interfaces) 11.6 BOD Analyst Pro With the aid of the BOD Analyst Pro software you can administrate BOD measurement on the PC and automatically calculate the measured values. More detailed information can be found in the BOD Analyst Pro operating manual.
Page 82
Transmitting data (USB interfaces) MultiLab 4010-1 ba76141e03 03/2014...
MultiLab 4010-1 Printer (only MultiLab 4010P-1) Printer (only MultiLab 4010P-1) 1 Operating LED – Illuminated: Printer is switched on – Flashing: Printer is switched off 2 Control panel with the keys <ON/FEED> and <OFF> 3 Lever to open the paper compartment 4 Lid of the paper compartment 12.1 Commissioning / switching the printer on or off Switching on the...
Printer (only MultiLab 4010P-1) MultiLab 4010-1 12.2 Operation / printing Data are only output to the printer if the following conditions are met Data are transmitted manually or automatically (see section 11.4 O PTIONS PC, page 80) FOR DATA TRANSMISSION TO A ...
MultiLab 4010-1 Printer (only MultiLab 4010P-1) 12.5 What to do if... / printer Cause Remedy Integrated printer – Printer switched off – Switch on the printer does not print (LED flashing) (LED illuminated) – No power pack connected – Connect the power pack –...
Maintenance, cleaning, disposal MultiLab 4010-1 Maintenance, cleaning, disposal 13.1 Maintenance 13.1.1 General maintenance activities The only maintenance activity required is replacing the batteries. See the relevant operating manuals of the IDS sensors for instruc- tions on maintenance. 13.1.2 Replacing the batteries You can operate the meter either with normal batteries or with rechargeable batteries (Ni-MH).
MultiLab 4010-1 Maintenance, cleaning, disposal Dispose of used batteries according to the local regulations of your country. End users within the European Union are obligated to return used batteries (even ecologically compatible ones) to a collection point set up for recycling purposes. Batteries are marked with the crossed-out waste container symbol.
What to do if... MultiLab 4010-1 What to do if... 14.1 pH More information and instructions on cleaning and exchange of sensors are given in the documentation of your sensor. Error message Cause Remedy OFL, UFL IDS pH sensor: – Measured value outside the –...
MultiLab 4010-1 What to do if... No stable measured Cause Remedy value IDS pH sensor: – pH sensor contaminated – Clean pH sensor Test sample: – pH value not stable – Measure with air excluded if neces- sary – Temperature not stable –...
What to do if... MultiLab 4010-1 Error message, Cause Remedy Error – Sensor contaminated – Clean the sensor – Measured temperature value – Keep to the temperature range for outside the operating conditions the test sample (display of OFL/UFL instead of a temperature value) –...
Page 91
MultiLab 4010-1 What to do if... Cause Remedy – Batteries almost empty – Replace the batteries (see section 13.1 M , page 86) AINTENANCE Display Instrument does not Cause Remedy react to keystroke – Operating condition undefined – Processor reset: or EMC load unallowed Press the <ENTER>...
Technical data MultiLab 4010-1 Technical data 15.1 Measuring ranges, resolution, accuracy Measuring ranges, Parameter Measuring range Accuracy accuracy Air pressure (abso- 225 ... 825 mm Hg ± 3 mm Hg lute)* Available only if a D.O. sensor is connected Further data are given in the documentation of your sensor. 15.2 General data Dimensions MultiLab 4010-1:...
Page 93
MultiLab 4010-1 Technical data USB interface (device) Type USB 1.1 USB-B (Device), PC Baud rate Adjustable: 1200, 2400, 4800, 9600, 19200 Baud Data bits Stop bits Parity None Handshake RTS/CTS Cable length Max. 3 m Applicable directives EC directive 2004/108/EC and standards EN 61326-1 EN 61000-3-2...
Firmware update MultiLab 4010-1 Firmware update 16.1 Firmware update for the meter MultiLab 4010-1 You can find available firmware update files for your meter on the Internet. With the "Firmware Update " program and a PC you can update the firmware of the MultiLab 4010-1 to the newest version.
MultiLab 4010-1 Firmware update 16.2 Firmware-Update for IDS Sensors With the "Firmware Update" program and a PC you can update the firmware of an IDS sensor to the newest version. You can find available firmware update files for your IDS sensor on the Internet. For updating, connect the IDS sensor to the MultiLab 4010-1, and the MultiLab 4010-1 to a PC.
Glossary MultiLab 4010-1 Glossary pH/ORP Asymmetry see zero point Electromotive force of The electromotive force U of the combination electrode is the measur- an electrode able electromotive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the combination electrode. Its dependency on the pH results in the electrode function, which is char- acterized by the parameters, slope and zero point.
Page 97
MultiLab 4010-1 Glossary Salinity The absolute salinity S of seawater corresponds to the relationship of the mass of dissolved salts to the mass of the solution (in g/kg). In practice, this dimension cannot be measured directly. Therefore, the practical salinity according to IOT is used for oceanographic monitor- ing.
Page 98
Glossary MultiLab 4010-1 Measured parameter The measured parameter is the physical dimension determined by measuring, e. g. pH, conductivity or D.O. concentration. Measured value The measured value is the special value of a measured parameter to be determined. It is given as a combination of the numerical value and unit (e.
Appendix MultiLab 4010-1 Appendix 19.1 Oxygen solubility table Solubility of oxygen in mg/L in water exposed to water-saturated air at 760 mm Hg pressure. Salinity = Measure of quantity of dissolved salts in water. Chlorinity = Measure of chloride content, by mass, of water. S(0/00) = 1.80655 x Chlorinity (0/00) Temp Chlorinity: 0...
MultiLab 4010-1 Contact Information Contact Information 20.1 Ordering & Technical Support Telephone: (800) 897-4151 (937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET Fax: (937) 767-1058 Email: environmental@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 Internet: www.ysi.com When placing an order please have the following information available:...
Page 106
Contact Information MultiLab 4010-1 ba76141e03 03/2014...
Page 108
150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation. For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241;...
Need help?
Do you have a question about the YSI MultiLab 4010P-1 and is the answer not in the manual?
Questions and answers