Rotate the device in the direction of the arrows on the
display by 360º (Figure 8).
rotating the device in the direction of the arrows by 360º
After rotating the device press the ENTER button. In this
operating mode, the orientation of the display will change by
180º, the device must also be turned by 180º, respectively. This
is done to increase the sensitivity of the device and is connected
with the location of the detector.
To recalibrate the gyroscope, hold down the ENTER
button for at least 2 seconds.
This window displays DER, the pulse count rate, and a pie
diagram of the probability of finding a source of gamma
radiation. When working in this mode, you need to continuously
make half-rotational movements with the device and focus on
the fullness of the sectors on the display (Figure 9).
Figure 9 – Search diagram display mode
Figure 8 – Gyroscope calibration:
26