8
Maintenance

Reinstall the Pump Head Assembly

Reinstall the Pump Head Assembly

When
When reassembling the pump
Tools required
p/n
5023-2524
5067-5688
G4220-20013
G4220-20015
Parts required
#
1
Damage of the pump drive
C AU T I O N
Starting the pump when the pump head is removed may damage the pump
drive.
Never start the pump when the pump head is removed.
1
If needed, apply a small amount of grease on the back
of the screws. Normally, the grease added during
manufacturing is sufficient for a long time.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Hex Key Set + Driver
A
Torque wrench 1 – 25 Nm with 14 mm wrench
A
4 mm hex bit
A
Adapter ¼ in square to hex
p/n
Description
A
79846-65501
Pump head grease
2
Slide the pump head assembly onto the pump drive.