3
Using the Pump
Solvent Information
Table 15
Part
Degasser chamber
MCGV
Active inlet valve
Outlet valve
Adapter
Pump head (body)
Pistons
Piston seals/wash seals
Pressure sensor
Purge valve
Damping unit
Capillaries/fittings
Tubings
Bio-inert Materials
For the Bio-inert LC system, Agilent Technologies uses highest-quality materials
in the flow path (also referred to as wetted parts), which are widely accepted by
life science scientists, as they are known for optimum inertness to biological
samples and ensure best compatibility with common samples and solvents over
a wide pH range. Explicitly, the complete flow path is free of stainless steel and
free of other alloys containing metals such as iron, nickel, cobalt, chromium,
molybdenum, or copper, which can interfere with biological samples. The flow
downstream of the sample introduction contains no metals whatsoever.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Materials in flow path (G5654A)
Materials
TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)
Platinum-iridium, titanium, PTFE, PEEK, FEP, PFA, ceramic, sapphire,
ruby
Platinum-iridium, titanium, gold, sapphire, ruby, ceramic, PTFE
Titanium, gold, ruby, ZrO
Titanium, gold
Titanium
Sapphire
PTFE, gold
Titanium
Titanium, gold, PTFE, ceramic
Titanium, gold
Titanium
PTFE
-based ceramic, tantalum
2
66