6
Error Information
Module Error Messages
Degasser: signal fail
Error ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser:
•
•
Probable cause
1
2
Degasser: Vacuum cannot be maintained
Error ID: 8014
Pressure in degasser vacuum chamber exceeds 180 mbar limit during normal
operation.
Probable cause
1
2
3
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
No valid pressure signal is shown during startup of the degasser
Measured pressure is higher than 32000 hPa
Degasser sensor defect.
Degasser sensor not connected to
mainboard.
Liquid in degasser tubing.
Leak in degasser tubing or chamber.
Degasser vacuum pump defective.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
109