Running The Test - Agilent Technologies InfinityLab LC Series User Manual

Isocratic and quaternary pumps
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Test Functions and Calibration
Leak Rate Test

Running the Test

Parts required
p/n
5067-6127
Running the test from the Agilent Lab Advisor
1 Select the Leak Rate Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the
NOTE
test has finished.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual