6
Error Information
Module Error Messages
Initialization Failed
Error ID: 2207, 2217
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable cause
1
2
3
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Blocked (passive or active) inlet valve.
Defective pump drive assembly.
Defective mainboard.
Suggested actions
Exchange the (passive or active) inlet valve.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
107