Exchange the Seal Wash Cartridge
Exchange the Seal Wash Cartridge
Parts required
p/n
5065-4445
Preparations
•
•
1
Remove the wash solvent tubings from the support ring
outlet and from the adapter leading to the waste bottle.
3
Put the new peristaltic pump cartridge onto the rod of
the pump motor and push the plastic clips into the
module housing.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Description
A
Peristaltic pump with PharMed tubing
Switch off pump at the main power switch.
Open the doors.
2
Unclip the peristaltic pump cartridge from the module
housing and remove it.
4
Connect the peristaltic pump tubes to the support rings
outlet and to the adapter leading to the waste bottle.