3
Using the Pump
Solvent Information
Table 16
Used Bio-inert materials
Module
Agilent 1260 Infinity II Bio-inert Pump
(G5654A)
Agilent 1260 Infinity II Bio-inert Multisampler
(G5668A)
Agilent 1260 Infinity II Bio-inert Manual Injector
(G5628A)
Agilent 1260 Infinity II Bio-inert Analytical Fraction Collector
(G5664B)
Bio-inert Flow Cells:
Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar ( 12 MPa) for MWD/DAD,
includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 –
PEEK fittings (G5615-60022)
(for Agilent 1260 Infinity II DAD G7115A, and MWD G7165A)
Bio-inert flow cell, 8 µL, 20 bar (pH 1–12), includes and PEEK fittings
(G5615-60005)
(for Agilent 1260 Infinity II FLD G7121A/B)
Bio-inert Heat Exchangers, Valves and Capillaries:
Quick Connect Heat Exchanger Bio-inert (G7116-60041)
(for Agilent 1260 Infinity II Multicolumn Thermostat G7116A)
Bio-inert Valve heads (G4235A, G5631A, G5632A, G5639A)
Bio-inert Connection capillaries
To ensure optimum biocompatibility of your Agilent 1260 Infinity II Bio-inert LC
NOTE
system, do not include non-inert standard modules or parts to the flow path. Do
not use any parts that are not labeled as Agilent "Bio-inert". For solvent
compatibility of these materials, see
Compatibility"
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
on page 68.
Materials
Titanium, gold, platinum-iridium, ceramic,
ruby, PTFE, PEEK
Upstream of sample introduction:
Titanium, gold, PTFE, PEEK, ceramic
Downstream of sample introduction:
PEEK, ceramic
PEEK, ceramic
PEEK, ceramic, PTFE
PEEK, ceramic, sapphire, PTFE
PEEK, fused silica, PTFE
PEEK (steel-cladded)
PEEK, ceramic (Al
2
Upstream of sample introduction:
Titanium
Downstream of sample introduction:
Agilent uses stainless-steel-cladded
PEEK capillaries, which keep the flow
path free of steel and provide pressure
stability up to 600 bar.
"General Information about Solvent/Material
O
based)
3
67

Advertisement