Exchange the Passive Inlet Valve
3
Insert the new valve into the pump head and tighten the
valve using a torque wrench (approx. 10 Nm).
approx.
10 Nm
5
Close the doors.
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
4
Reconnect the solvent inlet tube to the passive inlet
valve.
133