5.2.2 Define each Input Channel

1. Tap each Input channel button to set the options
2. Error type -the deviation from which the
3. Linearity - the transfer characteristic from
4. Tolerance - the test limits for the deviation from the
06 June 2013
that follow:
• Scaling - the scaling values are four set
values
• The maximum and minimum Reference
signal values (Reference High and Reference
Low)
• The Input signal values (Input High and Input
Low)
• The Input signals should relate to the
maximum and minimum Reference signal
values
transfer characteristic is to be calculated.
This can be one of the following:
• % Span as a percentage of the input signal
span
• % Rdg as a percentage of the input signal
reading
Reference to input signal. This can be one of the
following:
• Linear:
• Square Root:
transfer characteristic
a proportional response
often found in flow sensors