Table 8-6 millivolts (source) error limits
Source mV
06 June 2013
Calibrator
uncertainty
(mV)
0
100
200
1000
2000
Table 8-7 Voltage (source) error limits
Source
uncertainty
V
0
5
10
15
20
0.0001
0.00046
0.0009
0.003
0.006
Calibrator
(V)
0.000004
0.000019
0.000034
0.000049
0.000064
Druck DPI620 Genii
Permitted
DPI 620 error
(mV)
0.008
0.0125
0.017
0.1
0.14
Permitted
DPI 620 error
(V)
0.00042
0.0007
0.00010
0.00013
0.0016
Page 143 of 179