8.9

Procedures (CH1): True Ohms (measure)

Do the procedure as follows:
1. Repeat procedure 8.8; in step 3 and 4 select True Ohms
2. Check the error is within limits (Ref: Table 8-1)
Table 8-12 True Ohms (measure) error limits
Standard
Resistor
06 June 2013
uncertainty
(Ω)
0 (short
circuit)
100
200
300
400
1000
2000
4000
Resistor
DPI 620 error
(Ω)
-
0.002
0.004
0.006
0.008
0.02
0.04
0.08
Druck DPI620 Genii
Permitted
(Ω)
0.004
0.0095
0.015
0.0205
0.026
0.059
0.114
0.224