Procedures (Ch1): Tc Mv (Measure Or Source) - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
8.11

Procedures (CH1): TC mV (measure or source)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1)
2. Let the equipment get to a stable temperature
(minimum: 5 minutes `since the last power on)
3. Use the calibration menu (Ref: Section 8.1) to do a three-
point calibration
(-10 mV, Zero and 100 mV) for the measure or source
function
4. Check the calibration is correct:
• Select the applicable TC mV (measure) or (source)
function (ref Section 2.12)
• Apply the necessary values
• TC mV (measure): -10, 0 (short circuit)
• TC (mV): 10, 25, 50, 100
• TC mV (source): -10, 0, 10, 25, 50, 100
5. Check the error is within limits (Ref: Table 8-14)
Table 8-14 TC mV (measure or source) error limits
Input
or
output
TC (mV)
-10
0
10
25
50
100
06 June 2013
Calibrator uncertainty
TC (mV)
(measure)
0.0036
0.0036
0.0036
0.0036
0.0037
0.004
(source)
(measure)
0.00011
0.0001
0.00011
0.00015
0.00025
0.00046
Druck DPI620 Genii
Permitted
DPI 620 error (mV)
(source)
0.0085
0.0090
0.008
0.0085
0.0090
0.0091
0.0100
0.010
0.0125
0.0125
0.0170
Page 155 of 179