Note 1:
The DPI 620 has been assessed to the European
IEC60529 standard as having an ingress protection
rating of IP55, but this is for reliability purposes and not
for safety reasons
Note 2:
To meet the immunity requirements of annex A of
EN61326-1:2006, when used in an industrial
environment, the unit must be battery powered to
guarantee measurement specification
Note 3:
The case of the DPI620 is not suitable for prolonged
exposure to UV
Note 4:
The DPI620 is not suitable for permanent installation in
an outdoor environment