Preferences - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
6.13

Preferences

Select the applicable icon to change the poll address, short
tag or long tag
• Poll Address - when each transmitter has a unique
address
• Short tag - if the transmitter supports 8 character tags
• Long tag - if the transmitter supports 32 character
tags
Transmitters with a non-zero poll address default to a fixed
loop current of 4 mA. By default, the Druck DPI 620 Genii
polls address 0 (zero) only. Change the Poll Address by
selecting the appropriate search radio button or entering
the tag name in the search field
06 June 2013
Druck DPI620 Genii
Page 100 of 179