4. Check the calibration is correct:
• Select the Resistance (source) function (ref section
2.11)
• Apply the resistance values that are applicable to the
calibration (Ref: Table 8-13)
• Check the error is within limits (Ref: Table 8-13)
Table 8-13 Resistance (source) error limits
Ohms
(Ω)
0
100
200
300
400
1000
2000
3000
4000
06 June 2013
Excitation
uncertainty
(mA)
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
0.1
Calibrator
(Ω)
0.0014
0.0016
0.0021
0.0028
0.0035
0.008
0.016
0.024
0.032
Druck DPI620 Genii
Permitted
DPI 620
error (Ω)
0.014
0.038
0.062
0.086
0.11
0.31
0.55
0.86
1.1