Procedures (CH1): Resistance measure)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref:
Table 8-1)
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on)
3. Use the calibration menu (Ref: Section 8.1) to do a two-
point calibration
Range: 0-400Ω
• Nominal zero ohms:
Make a 4-wire connection to the 0Ω
• Nominal positive full-scale ohms
Make a 4-wire connection to the 400 Ω resistor
Range: 400Ω-4kΩ
• Nominal 400Ω
Make a 4-wire connection to the 400Ω resistor
• Nominal positive full-scale ohms
Make a 4-wire connection to the 4k resistor