Procedures (Ch1/Ch2): Current (Measure) - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
To do a calibration on a measure or source function, use the
advanced menu option.
1. Select:
2. Enter the calibration PIN: 4321
3. Select the
4. Select:
Then select a function and start the calibration:
1. Select channel
2. Select function
3. Select range (if applicable)
4. Follow on-screen instructions
When the calibration is complete, set the next calibration
date
8.2

Procedures (CH1/CH2): Current (measure)

Do the procedure as follows:
1. Connect the applicable calibration equipment (Ref: Table
8-1)
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on)
06 June 2013
DASHBOARD >>
button.
PERFORM CALIBRATION
ADVANCED
Druck DPI620 Genii
Page 136 of 179