User Manual
4. Check the calibration is correct
• Select the applicable AC mV or AC Volts (measure)
function (ref Section 0)
5. Apply the input values that are applicable to the
Calibration
• AC mV: 10, 500, 1000, 2000
• AC Volts: 5, 10, 20
6. Check the error is within limits (Ref: Table 8-15 or Table
8-16)
Table 8-15 AC mV (measure) error limits
Table 8-16 AC Volts (measure) error limits
Applied
AC Volts
06 June 2013
Applied
AC mV
uncertainty
10
500
1000
2000
Calibrator
uncertainty
(V)
5
10
20
Calibrator
(mV)
0.12
0.2
0.28
0.44
0.0018
0.0026
0.0042
Druck DPI620 Genii
Permitted
DPI 620 error
(mV)
2.5
3.1
3.75
5.0
Permitted
DPI 620 error
(V)
0.03
0.037
0.050
Page 160 of 179