4.7 Head support

Install the head support by following the steps below.
Take out the head support provided from the box.
Insert the head support to chair and secure with the screw provided on the backside of the chair.
The height of head support can also be adjusted with the same screw.

4.8 Connecting cables

Step 1:
Flip the chair base power switch to off, see Figure 10
Step 2:
Connect the Emergency Stop box to the base of the chair and place it on the desk or equipment cart for easy
operator access, see Figure 10
Step 3:
Connect the USB A-B cable from the base of the chair to the computer USB 3.2 HUB, see Figure 10
Step 4:
Connect the Orion Chair power cord from the chair base to the isolation transformer, see Figure 10
Step 5:
Connect the isolation transformer to wall power mains. The transformer output voltage is preset from the
factory, see Figure 10
Step 6:
Flip the chair base power switch to on, see Figure 10
Step 7:
Check that the power switch on the isolation transformer is set to on
Step 8:
Connect the goggle USB cables to the side of the Orion Chair Figure 10, when the system is powered on
and the Emergency Stop Box is lit up
Step 9:
Insert the goggle USB cables into the clips located on the back rest of the chair, see Figure 34
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Figure 33: Installing the head support
Page 30