With the drives set up, the installer can proceed to the registration of the Orion Chair in the VisualEyes
software.
3.2.1.5 Orion Chair registration in VisualEyes
There are two possible registration scenarios for the Orion Chair.
Scenario one (New complete VNG setup)
Installing a system ordered with new computer, VNG system, and accessories including Orion Chair.
This scenario will have licenses for both VNG system and Orion Chair installed from the factory, thus; the
installer can open the VisualEyes software entering testing directly. In this case the installer can now jump to
locking of the chair, see section 4.
Scenario two (Installing Orion Chair to existing VisualEyes system)
When new hardware is connected to an existing VisualEyes™ system it needs to be registered in the
software to be recognized correctly.
Step 1:
If a VisualEyes system is running with goggles and no Orion Chair, the installer must remove the goggles
currently used. This is done via System Configuration – Hardware and Licenses clicking Remove license
then click Unregister in the row mentioning the goggles, see figure 12 below.
Figure 15: Removing existing goggle licenses
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Page 18