Orion Reclining Chair
o
Isolation Transformer
Power Cord for isolation transformer
o
o
Power Cord for Orion Reclining chair 3m
o
USB A-B 3m
Emergency stop box
o
o
VNG goggle
3.2.1.1 Isolation Transformer
The installer must check the setting of the Isolation Transformer before it is plugged in the wall outlet.
The Transformer is set by factory, however; it is expected that the installer makes sure that the input voltage
for the Transformer is matching the mains power of the wall outlet (115V or 230V).
If not, use a flathead screwdriver to unlock the protective case. Using the same screwdriver, pry and slide out
the fuse sections. Rotate the module 180 degrees so that the right voltage setting is at the top. Slide the
module into place while pressing firmly with your finger. Secure the connector, see Figure 8.
Figure 8: Changing input voltage setting of the Isolation Transformer
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Page 14