Follow the steps mentioned below the achieve correct alignment:
Establish all connections between chair, PC, goggles, and safety transformer so the system is fully
functioning. Turn on the PC and enter the license codes as from section 3.2.1.
Place the chair in a position where the service door and connection panel are facing the center of the
TV.
Press the 'Home Chair" button in the software and check that the chair stops and locks with the
center of the chair seat aligned to the center of the TV.
Open the service chair panel to verify that the lock is fully engaged with the sprocket. If not fully
engaged, move the chair seat a few degrees by hand until the lock engages fully.
Adjust the chair base till the center of the chair seat and the TV center are aligned, see Figure 27. If
required, it is possible to create a different home position than the mechanical zero position. This is
done in the System Default Settings under the Rotary Chair tab.
Once you have determined the final position of the chair base, you can proceed with the drilling
process.

4.3 Drilling anchor holes

After careful assessment of the chair position, mark the area where the three holes are to be drilled.
You can either move the chair to the side after marking the positions for drilling or continue with drilling by
keeping the chair in place. Mark drill bit with electrical tape for hole depth of 1.75 to 2 inches (4.5 to 5 cm),
see Figure 28.
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Figure 27: Orion Chair center illustration
Figure 28: Drill bit with tape marking
Page 26