Interacoustics Micromedical Orion Installation Instructions Manual page 20

Reclining chair
Table of Contents

Advertisement

3.2.1.3 Connect chair to power and computer
Step 1:
Flip the chair base power switch to off, see Figure 10
Step 2:
Connect the Emergency Stop box to the base of the chair and place it on the desk or equipment cart for easy
operator access, see Figure 10
Step 3:
Connect the USB A-B cable from the base of the chair to the computer USB HUB, see Figure 10
Step 4:
Connect the Orion Chair power cord from the chair base to the isolation transformer, see Figure 10
Step 5:
Connect the isolation transformer to wall power mains. The transformer output voltage is preset from the
factory, see Figure 10
Step 6:
Flip the chair base power switch to on, see Figure 10
Step 7:
Check that the power switch on the isolation transformer is set to on
Step 8:
When the system is powered on and the Emergency Stop Box is lit up, connect the goggle USB cables to the
side of the Orion Chair Figure 10, and click them into the cable support clip next to the USB ports
Figure 10: Orion Reclining Chair connection diagram
The Orion Chair is now connected, and the installer can start installing drivers. Instruction hereof can be
found in the following sections.
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Page 16