Interacoustics Micromedical Orion Installation Instructions Manual page 26

Reclining chair
Table of Contents

Advertisement

To retrieve the settings for the Orion Chair, follow the path in Figure 21 from the VisualEyes home screen:
Figure 21: Path to Orion Reclining settings
Figure 22 below illustrates the Orion Chairs setting page, wherefrom modifications to chair performance can
be adjusted.
Figure 22: Orion Reclining settings page
3.2.2.1 Drift calibration
It is recommended to place a reference object close to the Orion Chair footrest (~1mm = 0,04inch) during
drift calibration, to tell if the Orion Chair is moving towards or away from the reference. The drift calibration
procedure is described in Figure 23.
Figure 23: Orion Reclining drift calibration process
D-0120794-D – 2023/06
Orion Reclining Chair Installation Instructions
Page 22