4.2.4
Power requirements
Introduction
This section describes the electrical power requirements for the Sepax C-Pro instru-
ment.
Electrical power requirements
The following table specifies the power requirements for the Sepax C-Pro instrument.
For power requirements for the computer, refer to the manufacturer's specifications.
Parameter
Supply voltage
Frequency
Power outlet type
Transient overvoltages
Power consumption
Location of sockets
Uninterruptible power supply (UPS)
Connecting the system to a UPS can prevent data loss during a power failure, and allow
time for a controlled shutdown of the instrument.
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
WARNING
Protective ground. The instrument must be connected to a
grounded power outlet.
WARNING
Access to power switch and power cord with plug. Do not
block access to the power switch and power cord. The power switch
must always be easy to access. The power cord with plug must
always be easy to disconnect.
Requirement
100 to 240 V~ autorange
50/60 Hz
Grounded mains outlets
Overvoltage category II
200 VA
Maximum 2 m from the instrument (due to length of
mains cable).