cytiva Sepax C-Pro Operating Instructions Manual page 153

Table of Contents

Advertisement

Error ID
Message
10227
Incorrect shutdown
10230
Device disconnected
10501
Default load setting
16010
Line pressure error
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
Description
Indicates incorrect shut-
down of the instrument
during the previous
procedure.
Communication problem
with the motor.
An incorrect shutdown
has happened and the
dynamic settings have
been corrupted. The
default ones are loaded.
Line pressure is incor-
rect.
8 Troubleshooting
8.2 Warning and error messages
Action
1. To avoid this
message, always use
QUIT in the starting
menu before
switching off the
instrument.
2. If the error message
appears in the middle
of the procedure,
contact Cytiva.
1. Shutdown the instru-
ment and resume
1
procedure
.
2. Contact Cytiva if the
error message
appears again.
Verify the settings and
traceability options of
the instrument and the
applications.
1. During instrument
start-up, verify that
the line pressure
sensor is free.
2. Make sure there are
no clots in the
product or kinks on
the lines.
3. Close all clamps.
4. Resume the proce-
1
dure
.
153