8.2
Warning and error messages
Error ID
Message
10101
Centrifuge blocked
10102
Centrifuge overspeed
10103
Centrifuge opened
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
Description
Centrifuge rotor locked.
Centrifuge speed was too
high.
Centrifuge covers were
opened during centrifu-
gation.
8 Troubleshooting
8.2 Warning and error messages
Action
1. Follow instructions
on screen.
2. Open the centrifuge
covers.
3. If the chamber can be
rotated, remove the
chamber and any
debris, and resume
the procedure.
4. If the chamber
cannot be rotated
and no debris is
found, take one of the
following actions:
• If available, use
another Sepax C-
Pro to resume the
1
procedure
.
• Otherwise, stop
using the instru-
ment and contact
Cytiva.
1. Close all the clamps.
2. Follow instructions
on screen.
3. If available, use
another Sepax C-Pro
instrument to resume
1
the procedure
.
4. Stop using the instru-
ment and contact
Cytiva.
1. Close all the clamps.
2. Follow instructions
on screen.
3. Resume procedure
1
.
146