Error ID
Message
10113
Check chamber pres-
sure
10114
Check line pressure
10116
Chamber underpres-
sure
10117
Line underpressure
10119
Chamber overpres-
sure
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
Description
Incorrect start-up
chamber pressure or
chamber pressure
sensor failure.
Incorrect start-up line
pressure or line pressure
sensor failure.
The vacuum under the
piston is too high.
Line underpressure.
The pressure under the
piston is too high.
8 Troubleshooting
8.2 Warning and error messages
Action
Remove the kit (if
installed) and restart the
instrument.
1. Remove the air filter
(if installed) and vali-
date the error
message.
2. If error message
persists, shut down
and restart the
instrument.
1. Close all clamps.
2. Follow instructions
on the screen.
3. Resume procedure
4. If error persists, save
the kit and contact
Cytiva.
1. Verify that clamps are
open and stopcocks
well positioned.
2. Make sure there are
no clots in the
product or kinks on
the lines.
3. Close all clamps.
4. Resume the proce-
1
dure
.
1. Close all clamps.
2. Follow instructions
on the screen.
3. Resume procedure
4. If error persists, save
the kit and contact
Cytiva.
1
.
1
.
148