Error ID
Message
10120
Line overpressure
10121
Piston is unstable
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
Description
Line overpressure.
Failure during piston
stabilization. Too much
air could have been aspi-
rated into the chamber.
8 Troubleshooting
8.2 Warning and error messages
Action
1. Verify that all relevant
clamps are open and
stopcocks well posi-
tioned.
2. Make sure there are
no clots in the
product or kinks on
the lines.
3. Close all clamps.
4. Resume the proce-
1
dure
.
1. Close all clamps.
2. Follow the on-screen
instructions.
3. Open the centrifuge
covers.
4. Verify that the sepa-
ration chamber is well
inserted into the
centrifuge pit.
5. Close the centrifuge
covers.
6. Resume the proce-
1
dure
.
7. If the error persists,
contact Cytiva.
149