Space requirements
The instrument can be installed on a standard laboratory bench or on a trolley with
wheels.
Prepare a clean working area on a stable laboratory bench or trolley that complies with
the specifications in the following table. There must be sufficient space to access the
rear panel and for ventilation purposes.
Parameter
Minimum bench area for operating the
instrument (W x D)
Load capacity
Sepax C-Pro Operating Instructions 29350570 AE OM-3074
WARNING
Access to power switch and power cord with plug. Do not
block access to the power switch and power cord. The power switch
must always be easy to access. The power cord with plug must
always be easy to disconnect.
CAUTION
Make sure that the system is placed on a stable, level bench with
adequate space for ventilation.
W
D
Specification
47 x 70 cm
980 N (100 kg) or higher
4 Installation
4.2 Site preparation
4.2.2 Room requirements